Optional: substitutes for Capacity when computing TP #3–#5.
Test Points
Test Point #1
—
—
Test Point #2
—
—
Test Point #3
—
—
Test Point #4
—
—
Test Point #5
—
—
Inputs
1/321/161/81/41/2
Optional: substitutes for Capacity when computing TP #3–#5.
Test Points & Tolerances
#
Test Point
NIST HB44 T6
NIST HB44 T1.1
USP
Recommended
—
HB44 T6 uses the class-specific breakpoints (in scale divisions) and is rounded to the nearest resolution; NIST/USP are floored to resolution and are at least 1×resolution.
Inputs
0.5110100
Tip: swapping flips “From” and “To”.
kggmgµglbozoz t
kggmgµglbozoz t
Result
—
—
Inputs
Result
Calculated Temperature
—
R₀
100 Ω
Resistance at −200 °C
18.520080 Ω
Resistance at 850 °C
390.481125 Ω
Inputs
Result
Calculated Resistance
—
R₀
100 Ω
Curve
PT-385
Temperature Range
−200 to 850 °C
RTD Conversion Notes
PT100 = 100 Ω at 0 °C; PT1000 = 1000 Ω at 0 °C.
PT-385 uses A = 3.9083×10⁻³, B = −5.775×10⁻⁷, C = −4.183×10⁻¹².
PT-392 uses A = 3.9787×10⁻³, B = −5.87×10⁻⁷, C = −4.0×10⁻¹².
Calculations use the Callendar–Van Dusen resistance-temperature equation.
This is a practical PT100/PT1000 RTD conversion aid, not a strict ITS-90 SPRT conversion. ITS-90 requires thermometer-specific calibration coefficients.
Use the sensor or instrument documentation/calibration certificate to select the proper curve.